Measurement of liquid′s specific heat capacity using changeable volume calorimeter

HE Mao-Gang,ZHONG Qiu,XUE Rong,ZHANG Ying
DOI: https://doi.org/10.3969/j.issn.1671-8097.2008.01.013
2008-01-01
Journal of Thermal Science and Technology
Abstract:Base on the steady-state theory,a non-flow calorimeter having a real-time changeable volume was designed and a specific heat capacity measurement system was constructed.In normal atmosphere,the specific heat capacities of pure water at temperatures ranging from 273.15~293.15 K were measured to verify the accuracy and reliability of the system.The measurement results show that the absolute average of relative deviations is 0.35% compared with the literature values.The correctness of the measuring principle and the accuracy and reliability of the experimental system were verified by the experimental results.At four different mixture ratios,specific heat capacities at constant pressure and specific heat capacities at constant volume for saturated liquid of oily R134a at temperatures ranging from 273.15~293.15 K were also measured.
What problem does this paper attempt to address?