Clinical application evaluation of Hitachi 7180 biochemistry autoanalyzer

姜春善,崔铉,金春花,任淑子
DOI: https://doi.org/10.3969/j.issn.1004-0412.2008.03.012
2008-01-01
Abstract:Objective To evaluate the application of Hitachi 7180 biochemistry autoanalyzer. Method Shengneng reagent and Shenneng control serum Ⅰ, Ⅱ(Lot NO.3229, Lot NO.4339) and patients' serum were used to evaluate the precision, linearity, accuracy and the schlepped polluting rate of 7180 biochemistry autoanalyzer by detecting these items as alanine aminotransferase (ALT), total protein (TP), calcium (Ca), glucose (GLU), creatine (Cre) and triglycerides (TG).Hitachi 7600 autoanalyzer was used as the control analyzer to evaluate the correlation. Results The within-day and day to day precision of Hitachi 7180 was both between 3% and 4%, the linearity was good (0.96~0.99), the schlepped polluting rate was 0, the coefficient of accuracy (CAC) was higher than 95% and compared with 7600 autoanalyzer, the coefficient of correlation was higher than 0.9856, with no significant difference. Conclusion The main index sign evaluation results of 7180 biochemistry autoanalyzer matches the design request, and fits all levels for biochemistry analysis.