Performance Evaluation of the Hitachi 7600-010 Biochemistry Autoaralyzer

许顺姬,姜春善,金幸
DOI: https://doi.org/10.3969/j.issn.1000-1824.2004.02.018
2004-01-01
Abstract:OBJECTIVE?To evaluate the performance of th e analyzer in order to access the utility of the Hitachi 7600-010 Biochemis try Autoanalyzer for the efficiency of routine chemistry work.?METHODS ? Precision, accuracy, linearity and recovery rate of 22 items withi n day and between day were evaluated. Hitachi 7150 was used as control analyzer to evaluate the correlation.?RESULTS?The CV almost all items w ere less than 3.0% except for D-BIL(7.71%), α-HBDH(6.34%). All items showe d good linearity in the perormance rance r0.99, slope 0.97~1.00. Hitac hi 7600-010 and Hitachi 7150 reveled good correlation (r0.981?8). The r ecovery rates for BUN, Cr, GLU, UA were in the range of 100.00%~102.73%.?CONCLUSI ON?Hitachi 7600-010 shows acceptable performance in the precision, lin earity and accuracy.
What problem does this paper attempt to address?