Evaluation on application of Roche biochemical analyzer for detection of glycosylated hemoglobin by enzymatic chemical method

JIANG Chun-shan,YOU Xin,JIN Chun-hua
2013-01-01
Abstract:OBJECTIVE To investigate the experimental results in determination of glycosylated hemoglobin(HbAlc)of enzymatic chemical methods by automatic biochemical analyzer.METHODS 50 cases of EDTA-K2 whole-blood from diabetes patients and healthy people were selected,and detected HbAlc by the Roche MODULAR PPI biochemical analyzer of.The degree of precision,range of linearity,anti-interference of the above method and IE-HPLC were compared.RESULTS The average coefficient of variation of the method was 0.78% and 1.69% within and between batches,HbAlc in 4.1%-14.0% range had a good linearity.The interference test showed the result had no significant interference in the assay(interference﹤5%,when vitamin C﹤500 mg/L,bilirubin﹤500 mg/L,hemolysin﹤5 000 mg/L).The correlation of enzymatic assay and HPLC was good(Y=1.033X-0.074 8,r=0.992 7).The expected bias was﹤1/2CLIA'88.CONCLUSION Enzymatic chemical method in determination of HbAlc has higher accuracy,precision and anti-interference,which is suitable for application in clinic.