Evaluation by NCCLS-EP9-A2 on consistency of two methods for glycosylated hemoglobin determination

WANG Yu-huan,JIANG Chun-shan,JIN Chun-hua
DOI: https://doi.org/10.3969/j.issn.1006-6187.2013.02.016
2013-01-01
Abstract:Objective To compare the results of HbA1c detected by the ion-exchange high-performance liquid chromatography(IE-HPLC) method and enzyme chemical method.Methods The experimental methodology mentioned in the document EP9-A2 of National Committee for Clinical Laboratory Standards(NCCLS) was used for comparison and analysis.Results The intra-and inter-assay coefficients of variation were all below 3%.In the regression analysis,the results of two groups were correlated,Y=1.033X-0.0748,r=0.9927,and the expected biases were 1.43% at 4%,2.53% at 10%,and 2.88% at 16% respectively.Conclusion The correlation and precision of HbA1c determination by IE-HPLC and enzymatic methods are fine and comparable.