The evaluation of Ketoamine oxidase method to determinate glycated albumin

JIA Ke-ke,LI Guo-quan,ZHANG Jie
2010-01-01
Abstract:Objective To evaluate Ketoamine oxidase method of determinating glycated albumin(GA).Methods To evaluate Lucica GA-L kit by studying the imprecision,dilution linearity,recovery,interference factors,and reagent stability.Analyze the correlation between GA and glycated hemoglobin(HbA1c);GA and fructosamine(GSP,NBT method).Compare bromocresol green method(BCG) with bromocresol purple method(BCP) of albumin determination.The results showed good precision,within-run CV <1.4%,between-run CV <2.4%;good linearity;the average recovery rate of 96.74% for GA,and the average recovery rate of 96.73% for ALB.Glucose,chyle and bilirubin had no significant interference on GA,but the hemoglobin affected the results of over 10% when the concentration was over 1.82 g/L.Results was stable for more than 4 weeks.It showed excellent correlation(r=0.894 0,P<0.001) between GA and HbA1c.Determination of albumin values with BCG method(±s,37.1 ±8.4 g/L) were higher than that of BCP method(±s,33.3±10.0 g/L),but with excellent correlation(r=0.990 8,P<0.001).When ALB concentration was over 30 g/L,GA was well correlated to GSP(r=0.847 4,P<0.001);When ALB concentration was less than 30 g/L,they were in poor correlation(r=0.569 7,P<0.05).Conclusion This method of GA determination shows good performance,and it can calculate the ratio of GA and albumin.There are fewer interference factors affecting this method.This method is a good index to monitor short-term average blood glucose levels.
What problem does this paper attempt to address?