Evaluation of immunosuppression turbidimetry method of determining glycosylated hemoglobin

姜春善,尤昕,金春花,金幸
DOI: https://doi.org/10.3969/j.issn.1006-6187.2010.01.006
2010-01-01
Abstract:Objective To evaluate immunosuppression turbidimetry method for deteriming the glycated hemoglobin by automatic biochemical analyzer.Methods 50 whole blood specimens from 25 diabetes patients and 25 healthy people were detected for HbA_1c by the Roche MODULAR PPI biochemical analyzer.We calculated the intra-and inter-coefficient of variation and the linear range of the research,and compard these with those by G7 glycosylated hemoglobin analyzer.Results The average intra-and inter-coefficients of variation were 2.0% and 3.7% for the method.HbA_1c was within 3.5%~ 21.0% range of good linearity.Compared with the glycated hemoglobin analyzer G7 method,the regression equation was y= 0.991x-0.06,r = 0.989.Conclusions Immunosuppression turbidimetry method for determining glycated hemoglobin by automatic analyzer has high accuracy and precision.It is suitable for routine clinical use.