Abstract:A new method using surface acoustic wave filter(SAWF) and electrostatic force in parallel-plate capacitor for expanding the measurement range of accelerometer with cantilever beam is put forward.The effect of electrostatic force on stabilizing the position of the sensitive mass block in the accelerometer and the elementary principle of expanding the measurement range using this effect are discussed.The calibration method for calibrating the accelerometer with cantilever beam is studied.The SAW accelerometer was calibrated using the relationship between the rotating speed of the centrifugal machine and its centripetal acceleration.