Carbonate‐Melt Oxidized Iridium Wire for Ph Sensing

M Wang,S Yao
DOI: https://doi.org/10.1002/elan.200302723
IF: 3.077
2003-01-01
Electroanalysis
Abstract:A new approach to the preparation of an iridium oxide film for pH sensing is demonstrated. A thick, uniform and dense ceramic oxide layer was grown on the surface of an iridium wire by oxidizing the wire in a molten alkali metal carbonate at high temperature. The alkali metal ion from the carbonate melt was incorporated into the oxide, resulting in a highly stable oxide compound. After treated in acid solution, the oxide layer became hydrated. SEM, XRD, TGA and element analysis were carried out to characterize the oxide film, and a possible formula of the oxide was computed as Li0.86IrO2.34(OH)(0.76) (.) 0.39H(2)O. The electrode made with this new type of oxide film exhibited good pH sensitivity and stability, even in strong acid/base solutions, or in strong corrosive solution such as hydrofluoric acid. Furthermore, the electrode showed excellent long-term stability over the test period of two and a half years. In addition, individual electrodes prepared from the same batch exhibited remarkable agreement with respect to potential/pH slopes and apparent standard electrode potentials. The performance of the electrodes depends on the properties of the oxide film, such as composition, hydration state and oxidation state.
What problem does this paper attempt to address?