Abstract:In this work, a new approach named "capillary melt method" was developed to fabricate micro antimony wires, and the wire surface was then oxidized in a nitrate melt at high temperature to obtain an antimony/antimony oxide pH electrode. Characterization results show that the oxide layer on the wire surface is porous, and consists of Sb2O3 crystal phase. The pH electrode, made using this method, showed good sensing performance in buffer solutions in the tested pH range of 2-12. Its EMF signal was found to have a linear relationship with pH value of the solution, with a sensitivity of 54.1 mV/pH and a fitting correlation coefficient of R-2 = 1.00. The advantages of the electrode are long-term stability, fast response, reproducibility and low cost.