Modeling and Accuracy Analysis of Hot Disk Thermal Constants Analyzer

王强,戴景民,张虎
DOI: https://doi.org/10.3969/j.issn.1004-1540.2008.04.004
2008-01-01
Abstract:The aim of the research presented in this paper is to enhance our understanding of the Hot disk thermal constants analyzer and to have an accurate evaluation of Hot disk thermal constants analyzer,the average temperature increase in the sensor surface was deduced by using Green's function method.Based on the simplified structure of the disk probe and the standardized material Pyroceram 9606,with a thermal conductivity of 4.05 W/(mK) at room temperature and under normal pressure was investigated and analyzed by using the data acquisition system of the Hot disk thermal constants analyzer and a new developed model.The calculation result shows that the model precision has a relative error of 6.17% estimated for thermal conductivity compared with the measurement result of the Hot disk,and the effects of the heat capacity of the Hot disk probe,the time delay of the data acquisition system and the elimination of heat from the four wire resistance in actual measurement are the main influencing factors of accuracy decrease.
What problem does this paper attempt to address?