Low Kilovotage 64-Slice Spiral CT in the Detection of Urinary Calculus:An Experimental Study

王秋霞,陈亮,黄文华,胡道予
DOI: https://doi.org/10.3969/j.issn.1000-0313.2011.08.002
2011-01-01
Abstract:Objective:To investigate the detectability of urinary calculi undergoing 64-slice CT scanning with reducing kV and to evaluate the image quality,so as to obtain the optimal scanning program.Methods:The calculi model was made by inserting 108 granules of urinary calculus containing four kinds of simple constituent into pork loin,and then scanned by 64-slice CT.The scanning parameters were as follows:three different dose-groups were chosen according to the fixed kV set 120kV,100kV,80kV;slice thickness and interval 5mm;the pitch is 0.984;the raw data were reconstructed into 1.25mm;all the reconstructed images were transferred to GE ADW4.3 workstation.To evaluate image quality according to different dose-groups and counted the number of detected calculus respectively.The standard deviation(SD) and signal to noise ratio(SNR) of the images were measured.CT dose index volume(CTDIvol) was obtained automatically.Results:120kV~250mAs was set to be the control group.Compared with the control group,80kV~150mAs,100kV~50mAs and 120kV~30mAs noise SD values were only increased by 7.60%,8.62% and 9.07%.The detection rate was 100%,while the radiation dose decreased significantly with the rate reaching 81.40%,87.60% and 87.98% respectively.Conclusion:It can reach 100% for the detection rate of urinary stones using the right combination of low kV-mAs scaning,and significantly reduce the radiation dose.