Calibration of copper-constantan thermocouple and related error analysis
Chen Liang,Hou Yu,Xi Lan
DOI: https://doi.org/10.3969/j.issn.1000-6516.2008.06.004
IF: 2.134
2008-01-01
Cryogenics
Abstract:Copper-constantan thermocouples were calibrated in the temperature range from –70 ℃ to 10 ℃ and error analysis was presented.A temperature measurement system which was based on Labview was developed.Effects of the welding methods and reference junctions on the accuracy of thermocouples were analyzed.Then the least square fittings with one order,second order,third order and fourth order were respectively performed on the test data.Error analysis on the calibration apparatus was conducted.The results show that thermocouples fabricated through soldering had comparative accuracy as thermocouples fabricated through butt welding;Model 7014 Multiplexer Card(Keithley Instruments,Inc.) and Aglient 34921A Multiplexer Module(Aglient Technologies,Inc.) were unsuitable for the temperature measurement at the reference junction in the calibration of thermocouples.
What problem does this paper attempt to address?