Feasibility study to apply EPID by trained therpistss in evaluation of set-up errors of pelvic irradiation

REN Jun,ZHANG Zhen,WU Xiao-ying,WANG Wen-chao,LU Hui-zhong,GUO Xiao-mei
DOI: https://doi.org/10.3969/j.issn.1007-3639.2006.12.012
2006-01-01
Abstract:Background and purpose:EPID is a tool which was conventionaly used for QA of radiotherapy.We investigated the feasibility of trained therapists use EPID to evaluation of set-up errors by analysis data difference between physicians and therapists.Methods:From May to Aug.2005,physicians and therapists collected two sets of 244 images separately from the same 12 patients.All patients got pelvic irradiation.Daily localization images were taken by EPID.Results:In the three directions of right to left,anterior to posterior and inferior to superior,the mean difference was(1.0±(0.9)) mm,(1.0±0.9) mm and(0.8±0.8) mm,respectively.There was no significant difference between physicians and therapists in EPID analysis.Conclusions:Trained therapists can use EPID analysis tools in evaluation set-up errors in pelvic irradiation.