The Proportional Ion Chamber and its Corresponding Circuit

Shao Beibei,Zhao Yan,Tang Yueli
2000-01-01
Abstract:High pressure electric current type ionization chambers have been used for pulsed radiation field measurement of accelerators for long time. The ion chamber like proportional tubes may improve the sensitivity and lower the cost of the ion chambers system. An amplify factor of 1,000 to 10,000 is introduced in the proportional area. With which the sensitivity of the tube is much higher than the ion chamber. At the ion chamber area, it fits the needs of very high radiation field. As the corresponding circuit, a low cost high sensitivity current digitizer has been designed and tested. It can respond to the weak current of 1 x 10 -13 A and has nice linearity between 5 x 10 -13 to 2 x 10-7 A. The system covers 108 order of radiation field. According to the level of radiation field, the high voltage on the tube is switched between proportional area and ion chamber area. A micro controller is employed for read out and the switch control. The system can be used for accelerators pulse radiation monitor and can also be used for beam loss measurement. 1 The detector