High-rate axial-field ionization chamber for particle identification of radioactive beams

J. Vadas,Varinderjit Singh,G. Visser,A. Alexander,S. Hudan,J. Huston,B. B. Wiggins,A. Chbihi,M. Famiano,M.M. Bischak,R. T. deSouza
DOI: https://doi.org/10.1016/j.nima.2016.08.056
2016-08-10
Abstract:The design, construction and performance characteristics of a simple axial-field ionization chamber suitable for identifying ions in a radioactive beam are presented. Optimized for use with low-energy radioactive beams (< 5 MeV/A) the detector presents only three 0.5 $\mu$m/cm$^2$ foils to the beam in addition to the detector gas. A fast charge sensitive amplifier (CSA) integrated into the detector design is also described. Coupling this fast CSA to the axial field ionization chamber produces an output pulse with a risetime of 60-70 ns and a fall time of 100 ns, making the detector capable of sustaining a relatively high rate. Tests with an $\alpha$ source establish the detector energy resolution as $\sim$8 $\%$ for an energy deposit of $\sim$3.5 MeV. The energy resolution with beams of 2.5 and 4.0 MeV/A $^{39}$K ions and the dependence of the energy resolution on beam intensity is measured. At an instantaneous rate of 3 x 10$^5$ ions/s the energy resolution has degraded to 14% with a pileup of 12%. The good energy resolution of this detector at rates up to 3 x 10$^5$ ions/s makes it an effective tool in the characterization of low-energy radioactive beams.
Instrumentation and Detectors,Nuclear Experiment
What problem does this paper attempt to address?
This paper aims to solve the problem of contaminant identification in radioactive ion beams (RIBs). Specifically, the paper presents the design, construction, and performance characteristics of a high - rate axial - field ionization chamber, which is suitable for particle identification in low - energy (<5 MeV/A) radioactive ion beams. To achieve this goal, the researchers developed a high - rate axial - field ionization chamber with a simple structure, named Rare Ion Purity Detector (RIPD), which is characterized by being able to operate at a high rate and maintain good energy resolution without affecting the ion - beam quality. By optimizing the ionization - chamber design and integrating a fast charge - sensitive amplifier (CSA), the RIPD can maintain an energy resolution of about 14% at an instantaneous rate of up to 3 x 10^5 ions/second, making it an effective tool for the characterization of low - energy radioactive ion beams. The paper experimentally tested the performance of the RIPD under different conditions, including measuring the energy resolution using an α - source and testing its performance at different beam - current intensities. The experimental results show that the RIPD can achieve an energy resolution of about 8% at low beam - current intensities, while at high beam - current intensities, due to the effects of recombination and signal pile - up, the energy resolution decreases but can still be maintained at around 14%, which is of great significance for the accurate measurement of low - energy radioactive ion beams.