Simple modeling of stress transfer at interface of cemented prosthesis and femur

YANG Zheng-ming,YU Xue-zhong,YAN Shi-gui,GUO Yi-mu
DOI: https://doi.org/10.3785/j.issn.1008-973X.2009.12.036
2009-01-01
Abstract:Assuming that the prosthesis,the cement and the femur are concentric cylinders made of linear elastic and isotropic materials,the axisymmetric interfacial simple stress transfer model considering thermal effect was established.When the prosthesis bears the compress stress,the distribution of axial stress of the prosthesis,the cement and the femur was analyzed by using the basic equations of axisymmetric elastic mechanics,and the interfacial shear stress at the prosthesis-cement interface and at the cement-femur interface in axial direction was obtained.Interfacial failure was the main failure form of prosthesis-cement-femur structure under the external load.The existence of thermal residual temperature difference accelerated the failure and destroy of the prosthesis-cement interface and the cement-femur interface,accelerated the relaxation of the prosthesis,and aggravated stress shielding effect of the femur.The main reason is that the cement has a relatively high thermal expansion coefficient.The elastic modulus and the thermal expansion coefficient of Ti-6Al-4V material were smaller than that of cobalt-chromium alloy material.The existence of thermal effect alleviated the failure of the interface and the femur to a certain degree,which is unanimous with the clinical finding.