The establishment and evaluation of white blood cell count for re-examination rules of automatic hematology analyzer

HE Chunyan,ZHANG Lei,WANG Jia,WENG Wenhao,LI Zhi
IF: 1.865
2011-01-01
Laboratory Medicine
Abstract:Objective To establish and validate re-examination optimized range of white blood cell(WBC) high and low values of automatic hematology analyzer Sysmex XE-2100(XE-2100).Methods 663 samples of WBC <4.0×109/L(low value group) and WBC>18×109/L(high value group) were enrolled from routine inpatients without hematology disease.The high value group was further classified into A:18.00-22.00×109/L(n=148),B:22.01-30.00×109/L(n=164) and C:>30.00×109/L(n=74).The low value group was classified into A':1.00-2.00×109/L(n=90),B':2.01-3.00×109/L(n=92) and C':3.01-4.00×109/L(n=95).According to the Sanitary Standard of People's Republic of China: the reference methods of WBC differential counting,the receiver operating characteristic(ROC) curve was used to find out an appropriate cut-off value for re-examination,and the value was validated.Results The area under the curve of A group was 0.820,the sensitivity was 1.0,and the specificity was 0.536 at the WBC value = 21.05×109/L.The areas under the curve of B and C groups were 0.749 and 0.329.The areas under the curve of A' and B' groups were 0.689 and 0.825.The sensitivity was 1.0,and the specificity was 0.317 at the WBC value = 2.45×109/L.The area under the curve of C' group was 0.983.The validating results showed that there were 28 samples without showing warning at WBC value = 2.45×109/L,among which 2 cases were false negative and basophilic cells>2 without clinical importance.7 samples showed warning,and one of them was false positive.There were 33 samples without showing warning at WBC value = 21.0×109/L,among which 1 was false negative and basophilic cells>2.7 samples showed warning,and 3 samples were false positive.Conclusions The warning system of XE-2100 provides useful information for re-examination.The high and low values of WBC counting are the factors for re-examination.The error of WBC classification by instrument is large,when WBC<2.50×109/L and WBC>21×10.009/L.Microscopic examination even without any warnings is needed.
What problem does this paper attempt to address?