Reliability Study of Leukocyte Classification by XE-2100 Automatic Hematology Analyzer

乐家新,王成彬,徐菡,宗佼,王红霞,朱红,练鑫
DOI: https://doi.org/10.3969/j.issn.1003-8868.2010.07.030
2010-01-01
Abstract:Objective To study the reliability of XE-2100 hematology analyzer in leukocyte classification.Methods The leukocyte classification by XE-2100 automatic hematology analyzer was compared with that by artificial classification.Results Leukocyte classification by XE-2100 automatic hematology analyzer showed a excellent precision when dealing with Neut,Lym,Mon,Eos and Baso.For the samples with normal WBC morphology,when the leukocyte classification result by XE-2100 automatic hematology analyzer compared with the artificial one,Neut,Lym and Eos took the best correlation(r20.96),and Mon took the second one(r2=0.848 60,and the correlation of Baso was poor(r2=0.497 5).The false positive rate was 26.02%,the false negative rate was 3.19%,the accuracy was 78.90%,the specificity was 73.98% and the sensitivity was 96.81%.The test results of the leukocyte sensitivity(alarm number)indicated that the alarm sensitivity of the blast cells was high in ANLL,and low in ALL.Conclusion There is a good precision of leukocyte classification results given by this hematology analyzer.It is accurate and reliable when WBC morphology of the samples is normal,and the main performance indicators are in line with the experimental requirements.For the samples which are suspected to be abnormal or alarmed,the leukocyte classification results must be reported after checking with the microscope.