Evaluation of BC-5500 hematology analyzer in analysis of leukocyte differential count

WENG Wenhao,LI Zhi,WANG Jia,LENG Jun,HUANG Bing,DONG Niuniu
IF: 1.865
2009-01-01
Laboratory Medicine
Abstract:Objective To evaluate the performance of domestic automatic hematology analyzer BC-5500,in leukocyte differential analysis.Methods The BC-5500 was evaluated according to the guidelines of International Council for Standardization in Haematology(ICSH) and National Committee for Clinical Laboratory Standards(NCCLS).Results BC-5500 achieved higher precision in detecting complete blood count,its total reproducibility,within group and between group coefficient variation(CV) in each parameter was from 0.27% to 2.85%.Stability was outstanding in 20 days for which its CV was from 0.57% to 2.93%.BC-5500 had an excellent linearity for white blood cell(WBC),red blood cell(RBC),hemoglobin(Hb) and platelet(PLT)(r>0.995)and minimal carryover.Compared with CELL-DYN 3700,both results were highly related.Evaluation of WBC differential count indicated an excellent correlation with microscopic examination for Neut and Lym(r=0.968 0,0.942 6 respectively),acceptable correlation for Mon and Eso(r= 0.646 5,0.820 7),but poor for Baso(r=0.287 6).In assessing the effect of the WBC warning by the BC-5500,alarm signals could indicate abnormal morphology cells in 172 cases.However its false positive rate was high and existed false negative to some extend.Compared with CELL-DYN 3700,the false negative of BC-5500 was a bit higher.Conclusions Automatic BC-5500 hematology analyzer is sensitive,accurate and rapid,but the improvement in abnormal morphology cell warnings is expected.As it is firstly made in our country and BC-5500 has its own reagent and quality control system,it can be greatly applicated in clinical detection of complete blood count and leukocyte differential.
What problem does this paper attempt to address?