Evaluation of sysmex XE-2100 hematology analyzer by using CLSI documents

Ma Xuebin,Yang Ming,Jia Jingyuan,Huang Xinqiang,Zhao Qiangyuan,Ma Cong
DOI: https://doi.org/10.3969/j.issn.1673-4130.2013.18.031
2013-01-01
Abstract:Objective To evaluate the performance of Sysmex XE-2100 Hematology Analyzer.Methods The evaluation was performed by using guideline established by the International Committee for Standardization Hematology(ICSH).Precision,linearity,carryover and the contrast were analyzed.Results The background counts were lower than the design.Precision,linearity,carryover of the WBC,RBC,HGB,HCT,PLT were good,the carryover was low,the results were different for Leukocyte Differential Count.Conclusion The performance of Sysmex XE-2100hematology analyzer is adequate for clinical demand,so it can be used for the blood samples analysis,and the microscopic reexamination is necessary.