Non-contact XUV metrology of Ru/B4C multilayer optics by means of Hartmann wavefront analysis

Mabel Ruiz-Lopez,Hugo Dacasa,Benoit Mahieu,Magali Lozano,Philippe Zeitoun,Davide Bleiner
DOI: https://doi.org/10.1364/ao.57.001315
2018-06-22
Abstract:Short-wavelength imaging, spectroscopy, and lithography scale down the characteristic length-scale to nano meters. This poses tight constraints on the optics finishing tolerances, which is often difficult to characterize. Indeed, even a tiny surface defect degrades the reflectivity and spatial projection of such optics. In this study, we demonstrate experimentally that a Hartmann wavefront sensor for extreme ultraviolet (XUV) wavelengths is an effective non-contact analytical method for inspecting the surface of multilayer optics. The experiment was carried out in a tabletop laboratory using a high-order harmonic generation as an XUV source. The wavefront sensor was used to measure the wavefront errors after the reflection of the XUV beam on a spherical Ru/B4C multilayer mirror, scanning a large surface of approximately 40 mm in diameter. The results showed that the technique detects the aberrations in the nanometer range.
Instrumentation and Detectors,Optics
What problem does this paper attempt to address?