Characterizing IMARAD CZT Detectors with Time Resolved Anode and Cathode Signals

J. S. Perkins,H. Krawczynski,P. Dowkontt
DOI: https://doi.org/10.48550/arXiv.astro-ph/0310910
2003-11-01
Abstract:We present the results of using standard IMARAD CZT detectors with a 100 MHz readout of the anode and cathode pulses. The detectors, 2 cm x 2 cm large and 0.5 cm thick, have 64 Indium pixellated anode contacts at a pitch of 2.5 mm. We investigate the possibilities to improve on the detector's photo-peak efficiency and energy resolution using two depth of interaction (DOI) indicators: (i) the total charge induced on the cathode and (ii) the drift time of the electron cloud determined from the anode and/or the cathode pulses. The DOI correction with the cathode charge gives better results, increasing the 662 keV photo-peak efficiency by 57% and improving the energy resolution from 2.33% to 2.15% (FWHM, including the electronic noise). The information on the time dependence of the induced charge can be used as a diagnostic tool to understand the performance of the detector. Detailed comparison of the pulse shapes with detector simulations gives the electron mobility and drift time and makes it possible to assess the weighting potential and electric field inside the detector
Astrophysics
What problem does this paper attempt to address?
This paper aims to solve the problem of performance optimization of CdZnTe (CZT) detectors in high - energy photon detection, especially to improve their photo - peak efficiency and energy resolution. A standard IMARAD CZT detector was used in the study, and time - resolved data of anode and cathode pulse signals were obtained at a 100 MHz readout rate. The authors explored two Depth of Interaction (DOI) indicators to improve the performance of the detector: 1. **Total Charge**: The total amount of charge induced on the cathode. 2. **Drift Time**: The drift time of the electron cloud determined from the anode and/or cathode pulses. It was found that using cathode charge as a DOI correction method is more effective than using drift time. Specifically, this correction method can increase the efficiency of the 662 keV photo - peak by 57% and improve the energy resolution from 2.33% to 2.15% (full width at half maximum, FWHM, including electronic noise). In addition, by analyzing the comparison between the pulse shape and the detector simulation results, the researchers were able to evaluate the weighting potential and electric field distribution inside the detector, thereby further understanding the working performance and optimization direction of the detector. These improvements are of great significance for the future applications of space - borne X - ray and gamma - ray telescopes, such as the EXIST (Energetic X - Ray Imaging Survey Telescope) and ACT (Advanced Compton Telescope) projects.