Background Measurements from Balloon-Borne CZT Detectors

Johnathan A Jenkins,Tomohiko Narita,Jonathan E. Grindlay,Peter F. Bloser,Carl Stahle,Brad Parker,Scott Barthelmy
DOI: https://doi.org/10.1117/12.461537
2002-12-04
Abstract:We report detector characteristics and background measurements from two prototype imaging CZT detectors flown on a scientific balloon payload in May 2001. The detectors are both platinum-contact 10mm x 10mm x 5mm CZT crystals, each with a 4 $\times$ 4 array of pixels tiling the anode. One is made from IMARAD horizontal Bridgman CZT, the other from eV Products high-pressure Bridgman material. Both detectors were mounted side-by-side in a flip-chip configuration and read out by a 32-channel IDE VA/TA ASIC preamp/shaper. We enclosed the detectors in the same 40deg field-of-view collimator (comprisinga graded passive shield and plastic scintillator) used in our previously-reported September 2000 flight. I-V curves for the detectors are diode-like, and we find that the platinum contacts adhere significantly better to the CZT surfaces than gold to previous detectors. The detectors and instrumentation performed well in a 20-hour balloon flight on 23/24 May 2001. Although we discovered a significant instrumental background component in flight, it was possible to measure and subtract this component from the spectra. The resulting IMARAD detector background spectrum (from 30 keV to ~450 keV) reaches ~5 x 10^{-3}$ counts/cm^2 -sec-keV at 100 keV and has a power-law index of ~2 at high energies. The eV Products detector has a similar spectrum, although there is more uncertainty in the energy scale because of calibration complications.