Measurements of time resolution of the RD50-MPW2 DMAPS prototype using TCT and 90 Sr

J. Debevc,M. Franks,B. Hiti,U. Kraemer,G. Kramberger,I. Mandić,R. Marco-Hernández,D.J.L. Nobels,S. Powell,J. Sonneveld,H. Steininger,C. Tsolanta,E. Vilella,C. Zhang
DOI: https://doi.org/10.1088/1748-0221/19/05/p05068
2024-05-24
Journal of Instrumentation
Abstract:Results in this paper present an in-depth study of time resolution for active pixels of the RD50-MPW2 prototype CMOS particle detector. Measurement techniques employed include Backside- and Edge-TCT configurations, in addition to electrons from a 90 Sr source. A sample irradiated to 5 · 10 14 n eq /cm 2 was used to study the effect of radiation damage. Timing performance was evaluated for the entire pixel matrix and with positional sensitivity within individual pixels as a function of the deposited charge. Time resolution obtained with TCT is seen to be uniform throughout the pixel's central region with approx. 220 ps at 12 ke - of deposited charge, degrading at the edges and lower values of deposited charge. 90 Sr measurements show a slightly worse time resolution as a result of delayed events coming from the peripheral areas of the pixel.