Traceable Calibration of Atmospheric Oxidized Mercury Measurements

Tyler R. Elgiar,Seth N. Lyman,Teodor D. Andron,Lynne Gratz,A. Gannet Hallar,Milena Horvat,Sreekanth Vijayakumaran Nair,Trevor O’Neil,Rainer Volkamer,Igor Živković
DOI: https://doi.org/10.1021/acs.est.4c02209
IF: 11.4
2024-06-09
Environmental Science & Technology
Abstract:Most previous measurements of oxidized mercury were collected using a method now known to be biased low. In this study, a dual-channel system with an oxidized mercury detection limit of 6-12 pg m^(-3) was deployed alongside a permeation tube-based automated calibrator at a mountain top site in Steamboat Springs Colorado, USA, in 2021 and 2022. Permeation tubes containing elemental mercury and mercury halides were characterized via an International System of Units (SI)-traceable gravimetric...
environmental sciences,engineering, environmental
What problem does this paper attempt to address?