Design and Application of On-line Atmospheric Mercury Analyzer

YU Zhongying,ZENG Limin,ZHANG Yuanhang
DOI: https://doi.org/10.3321/j.issn:0479-8023.2006.06.011
2006-01-01
Abstract:Based on the method of gold amalgamation and CVAFS, the authors designed an instrument for continuously measuring total gaseous mercury in ambient air. The instrument operates with purified ambient air as carrier gas and does not require Argon or Nitrogen for detection. The sampling is run at about 1 L/min, with sampling times of 10 min. Under these conditions, a detection limit of about 0.2 ng/m3 is achieved. It has been showed that the relative standard deviations for 0.1 ng mercury vapor samples is less than 1.4%, and the linearity of calibration curve is quite well with the correlation coefficient of 0.999 5. The instrument was used to continuously measure total gaseous mercury in the atmosphere at Peking University for about three weeks, observing an average concentration of TGM about 7.8 ng/m3.