Application and problem discussion of on-line atmospheric mercury analyzer in field measurement: take measurement in Wuxi in winter for example

Zhijun Chen,Yinsong Zhang,Limin Zeng,Yongjie Wei,XiaoYan Tang
2012-01-01
Abstract:During Dec. 30, 2010 to Jan. 14, 2011, measurement of total gaseous mercury (TGM) concentrations was carried out by using a set of online atmospheric mercury analyzer developed by Zeng's group at Caozhang station (E120° 17'40″, N31° 33'36″) in Wuxi, and the analyzer performanced well. The mean concentration over the sampling period was 7.37 ng/m 3 with a range from 2.33 to 66.85 ng/m 3. Some other elements in the ambient air could also be detected by the analyzer. Zinc which was probably released during metal smelting with mercury could be the reason, and this was provided by data analysis and lab experiment. It provides a way for promotion of the analyzer and makes the analyzer monitoring the particular source (smelting) of TGM possible after some improve-ments.