Dynamic X-ray imaging with screen-printed perovskite CMOS array

Yanliang Liu,Chaosong Gao,Dong Li,Xin Zhang,Jiongtao Zhu,Meng Wu,Wenjun Liu,Tongyu Shi,Xingchen He,Jiahong Wang,Hao Huang,Zonghai Sheng,Dong Liang,Xue-Feng Yu,Hairong Zheng,Xiangming Sun,Yongshuai Ge
DOI: https://doi.org/10.1038/s41467-024-45871-2
IF: 16.6
2024-02-21
Nature Communications
Abstract:Abstract High performance X-ray detector with ultra-high spatial and temporal resolution are crucial for biomedical imaging. This study reports a dynamic direct-conversion CMOS X-ray detector assembled with screen-printed CsPbBr 3 , whose mobility-lifetime product is 5.2 × 10 −4 cm 2 V –1 and X-ray sensitivity is 1.6 × 10 4 μC Gy air –1 cm –2 . Samples larger than 5 cm $$\times$$ × 10 cm can be rapidly imaged by scanning this detector at a speed of 300 frames per second along the vertical and horizontal directions. In comparison to traditional indirect-conversion CMOS X-ray detector, this perovskite CMOS detector offers high spatial resolution (5.0 lp mm −1 ) X-ray radiographic imaging capability at low radiation dose (260 nGy). Moreover, 3D tomographic images of a biological specimen are also successfully reconstructed. These results highlight the perovskite CMOS detector’s potential in high-resolution, large-area, low-dose dynamic biomedical X-ray and CT imaging, as well as in non-destructive X-ray testing and security scanning.