Stable perovskite single-crystal X-ray imaging detectors with single-photon sensitivity

Kostiantyn Sakhatskyi,Bekir Turedi,Gebhard J. Matt,Erfu Wu,Anastasiia Sakhatska,Vitalii Bartosh,Muhammad Naufal Lintangpradipto,Rounak Naphade,Ivan Shorubalko,Omar F. Mohammed,Sergii Yakunin,Osman M. Bakr,Maksym V. Kovalenko
DOI: https://doi.org/10.1038/s41566-023-01207-y
IF: 35
2023-05-09
Nature Photonics
Abstract:A major thrust of medical X-ray imaging is to minimize the X-ray dose acquired by the patient, down to single-photon sensitivity. Such characteristics have been demonstrated with only a few direct-detection semiconductor materials such as CdTe and Si; nonetheless, their industrial deployment in medical diagnostics is still impeded by elaborate and costly fabrication processes. Hybrid lead halide perovskites can be a viable alternative owing to their facile solution growth. However, hybrid perovskites are unstable under high-field biasing in X-ray detectors, owing to structural lability and mixed electronic–ionic conductivity. Here we show that both single-photon-counting and long-term stable performance of perovskite X-ray detectors are attained in the photovoltaic mode of operation at zero-voltage bias, employing thick and uniform methylammonium lead iodide single-crystal films (up to 300 μm) and solution directly grown on hole-transporting electrodes. The operational device stability exceeded one year. Detection efficiency of 88% and noise-equivalent dose of 90 pGy air are obtained with 18 keV X-rays, allowing single-photon-sensitive, low-dose and energy-resolved X-ray imaging. Array detectors demonstrate high spatial resolution up to 11 lp mm −1 . These findings pave the path for the implementation of hybrid perovskites in low-cost, low-dose commercial detector arrays for X-ray imaging.
optics,physics, applied
What problem does this paper attempt to address?