Cs2AgBiBr6 single-crystal X-ray detectors with a low detection limit

Weicheng Pan,Haodi Wu,Jiajun Luo,Zhenzhou Deng,Cong Ge,Chao Chen,Xiaowei Jiang,Wan-Jian Yin,Guangda Niu,Lujun Zhu,Lixiao Yin,Ying Zhou,Qingguo Xie,Xiaoxing Ke,Manling Sui,Jiang Tang
DOI: https://doi.org/10.1038/s41566-017-0012-4
IF: 35
2017-10-02
Nature Photonics
Abstract:Sensitive X-ray detection is crucial for medical diagnosis, industrial inspection and scientific research. The recently described hybrid lead halide perovskites have demonstrated low-cost fabrication and outstanding performance for direct X-ray detection, but they all contain toxic Pb in a soluble form. Here, we report sensitive X-ray detectors using solution-processed double perovskite Cs2AgBiBr6 single crystals. Through thermal annealing and surface treatment, we largely eliminate Ag+/Bi3+ disordering and improve the crystal resistivity, resulting in a detector with a minimum detectable dose rate as low as 59.7 nGyair s−1, comparable to the latest record of 0.036 μGyair s−1 using CH3NH3PbBr3 single crystals. Suppressed ion migration in Cs2AgBiBr6 permits relatively large external bias, guaranteeing efficient charge collection without a substantial increase in noise current and thus enabling the low detection limit.
optics,physics, applied
What problem does this paper attempt to address?