Initial Characterization of the SAFIR Prototype PET-MR Scanner

Christian Ritzer,Robert Becker,Alfred Buck,Volker Commichau,Jan Debus,Lubomir Djambazov,Afroditi Eleftheriou,Jannis Fischer,Peter Fischer,Mikiko Ito,Parisa Khateri,Werner Lustermann,Michael Ritzert,Ulf Röser,Markus Rudin,Ilaria Sacco,Charalampos Tsoumpas,Geoffrey Warnock,Matthias Wyss,Agnieszka Zagozdzinska-Bochenek,Bruno Weber,Günther Dissertori,Ulf Roser,Gunther Dissertori
DOI: https://doi.org/10.1109/trpms.2020.2980072
2020-09-01
IEEE Transactions on Radiation and Plasma Medical Sciences
Abstract:The SAFIR collaboration is currently developing a high-rate positron emission tomography (PET) insert to study fast kinetic processes in small animals. Our insert is designed for simultaneous image acquisition with a preclinical 7 T magnetic resonance (MR) imaging device. In contrast to existing preclinical PET scanners and inserts, our hardware is optimized for high-rate measurements with source activities up to 500 MBq. As a first step, the SAFIR Prototype insert was constructed. This already incorporates the final components, but has a reduced axial field-of-view (35.6 mm). We use lutetium-yttrium oxyorthosilicate crystals (2.12 mm $\times $ 2.12 mm $\times $ 13 mm) one-to-one coupled to silicon photomultipliers. All analog signals are digitized within the insert. We use 49 MR-compatible dc–dc converters in the insert to provide the power to all readout electronics. After shimming, no degradation of the homogeneity of the static $B_{0}$ field in the MR scanner was observed. During full operation, we saw a minor reduction in the signal-to-noise ratio of the MR data of 4.9%. With a low activity point source (22Na 0.65 MBq) we obtained a coincidence energy resolution of 13.8% full width at half maximum (FWHM) and a coincidence timing resolution of 194 ps (FWHM). First PET/MR rat brain and high-rate mouse cardiac images (84.9 MBq) are shown in this article.
What problem does this paper attempt to address?