NEMA characterization of the SAFIR prototype PET insert

Parisa Khateri,Werner Lustermann,Christian Ritzer,Charalampos Tsoumpas,Günther Dissertori
DOI: https://doi.org/10.1186/s40658-022-00454-2
2022-06-14
EJNMMI Physics
Abstract:The SAFIR prototype insert is a preclinical Positron Emission Tomography (PET) scanner built to acquire dynamic images simultaneously with a 7 T Bruker Magnetic Resonance Imaging (MRI) scanner. The insert is designed to perform with an excellent coincidence resolving time of 194 ps Full Width Half Maximum (FWHM) and an energy resolution of 13.8% FWHM. These properties enable it to acquire precise quantitative images at activities as high as 500 MBq suitable for studying fast biological processes within short time frames (< 5 s). In this study, the performance of the SAFIR prototype insert is evaluated according to the NEMA NU 4-2008 standard while the insert is inside the MRI without acquiring MRI data.
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?