PETcoil: first results from a second-generation RF-penetrable TOF-PET brain insert for simultaneous PET/MRI
Qian Dong,Muhammad Nasir Ullah,Derek R Innes,Ronald D Watkins,Chenming Chang,Sarah Jin Zou,Andrew Groll,Ilaria Sacco,Garry Chinn,Craig S Levin
DOI: https://doi.org/10.1088/1361-6560/ad7221
IF: 3.5
2024-08-24
Physics in Medicine and Biology
Abstract:Simultaneous PET/MRI provides concurrent information about anatomic, functional, and molecular changes in disease. We are developing a second generation MR-compatible RF-penetrable TOF-PET insert. The insert has a smaller scintillation crystal size and ring diameter compared to clinical whole-body PET scanners, resulting in higher spatial resolution and sensitivity. This paper reports the initial system performance of this full-ring PET insert. The global photopeak energy resolution and global coincidence time resolution, 11.74 ± 0.03 % FWHM and 238.1 ± 0.5 ps FWHM, respectively, are preserved as we scaled up the system to a full ring comprising 12,288 LYSO-SiPM channels. Throughout a ten-hour experiment, the system performance remained stable, exhibiting a less than 1% change in all measured parameters. In a resolution phantom study, the system successfully resolved all 2.8 mm diameter rods, achieving an average VPR of 0.28 ± 0.08 without TOF and 0.24 ± 0.07 with TOF applied. Moreover, the implementation of TOF in the Hoffman phantom study also enhanced image quality. Initial MR compatibility studies of the full PET ring were performed with it unpowered as a milestone to focus on looking for material and geometry-related artifacts. During all MR studies, the MR body coil functioned as both the transmit and receive coil, and no observable artifacts were detected. As expected, using the body coil also as the RF receiver, MR image signal-to-noise ratio exhibited degradation (∼30%), so we are developing a high quality receive-only coil that resides inside the PET ring.
engineering, biomedical,radiology, nuclear medicine & medical imaging