Demonstration of in-vivo simultaneous 3D imaging with 18F-FDG and Na131I using Compton–PET system
Donghwan Kim,Linlin Yan,Kenji Shimazoe,Hiroyuki Takahashi,Kenichiro Ogane,Masao Yoshino,Kei Kamada,Mizuki Uenomachi
DOI: https://doi.org/10.1038/s41598-024-71750-3
IF: 4.6
2024-09-11
Scientific Reports
Abstract:Simultaneous imaging of the SPECT tracer 131 I and PET tracer 18 F is important in the diagnosis of high- and low-grade thyroid cancers because high-grade thyroid cancers have high 18 F-FDG and low 131 I uptake, while low-grade thyroid cancers have high 131 I and low 18 F-FDG uptake. In this study, Na 131 I and 18 F-FDG were simultaneously imaged using the Compton-PET system, in vivo. The angular resolution and sensitivity of the Compton camera with 356 keV gamma ray measured using a 133 Ba point source were 12.3° and 2 × 10 −5 , respectively. The spatial resolution and sensitivity of PET were measured with a 22 Na point source. The transaxial and axial spatial resolutions of the PET at the center of the FOV were 1.15 mm and 2.04 mm, respectively. Its sensitivity was 1.2 × 10 −4 . In-vivo images of the 18 F and 131 I isotopes were simultaneously acquired from mice. These showed that 18 F-FDG was active in the heart, brown fat, and brain, while Na 131 I was active in the thyroid, stomach, and bladder. Artifacts were found in the Compton camera images when the activity of 131 I was much lower than that of 18 F. This study demonstrates the potential of simultaneous clinical imaging of 18 F and 131 I.