Walk-through flat panel total-body PET: a patient-centered design for high throughput imaging at lower cost using DOI-capable high-resolution monolithic detectors

Stefaan Vandenberghe,Florence M. Muller,Nadia Withofs,Meysam Dadgar,Jens Maebe,Boris Vervenne,Maya Abi Akl,Song Xue,Kuangyu Shi,Giancarlo Sportelli,Nicola Belcari,Roland Hustinx,Christian Vanhove,Joel S. Karp,Vandenberghe, Stefaan
DOI: https://doi.org/10.1007/s00259-023-06341-x
2023-07-20
Abstract:Long axial field-of-view (LAFOV) systems have a much higher sensitivity than standard axial field-of-view (SAFOV) PET systems for imaging the torso or full body, which allows faster and/or lower dose imaging. Despite its very high sensitivity, current total-body PET (TB-PET) throughput is limited by patient handling (positioning on the bed) and often a shortage of available personnel. This factor, combined with high system costs, makes it hard to justify the implementation of these systems for many academic and nearly all routine nuclear medicine departments. We, therefore, propose a novel, cost-effective, dual flat panel TB-PET system for patients in upright standing positions to avoid the time-consuming positioning on a PET-CT table; the walk-through (WT) TB-PET. We describe a patient-centered, flat panel PET design that offers very efficient patient throughput and uses monolithic detectors (with BGO or LYSO) with depth-of-interaction (DOI) capabilities and high intrinsic spatial resolution. We compare system sensitivity, component costs, and patient throughput of the proposed WT-TB-PET to a SAFOV (= 26 cm) and a LAFOV (= 106 cm) LSO PET systems.
What problem does this paper attempt to address?