Feasibility of the J-PET to monitor the range of therapeutic proton beams
Jakub Baran,Damian Borys,Karol Brzeziński,Jan Gajewski,Michał Silarski,Neha Chug,Aurélien Coussat,Eryk Czerwiński,Meysam Dadgar,Kamil Dulski,Kavya V Eliyan,Aleksander Gajos,Krzysztof Kacprzak,Łukasz Kapłon,Konrad Klimaszewski,Paweł Konieczka,Renata Kopeć,Grzegorz Korcyl,Tomasz Kozik,Wojciech Krzemień,Deepak Kumar,Antony J Lomax,Keegan McNamara,Szymon Niedźwiecki,Paweł Olko,Dominik Panek,Szymon Parzych,Elena Perez Del Rio,Lech Raczyński,Moyo Simbarashe,Sushil Sharma,Shivani,Roman Y Shopa,Tomasz Skóra,Magdalena Skurzok,Paulina Stasica,Ewa Ł Stępień,Keyvan Tayefi,Faranak Tayefi,Damien C Weber,Carla Winterhalter,Wojciech Wiślicki,Paweł Moskal,Antoni Ruciński,Kavya V. Eliyan,Antony J. Lomax,Elena Perez del Rio,Roman Y. Shopa,Ewa Ł. Stępień,Damien C. Weber
DOI: https://doi.org/10.1016/j.ejmp.2024.103301
IF: 3.119
2024-02-01
Physica Medica
Abstract:PURPOSE: The aim of this work is to investigate the feasibility of the Jagiellonian Positron Emission Tomography (J-PET) scanner for intra-treatment proton beam range monitoring.METHODS: The Monte Carlo simulation studies with GATE and PET image reconstruction with CASToR were performed in order to compare six J-PET scanner geometries. We simulated proton irradiation of a PMMA phantom with a Single Pencil Beam (SPB) and Spread-Out Bragg Peak (SOBP) of various ranges. The sensitivity and precision of each scanner were calculated, and considering the setup's cost-effectiveness, we indicated potentially optimal geometries for the J-PET scanner prototype dedicated to the proton beam range assessment.RESULTS: The investigations indicate that the double-layer cylindrical and triple-layer double-head configurations are the most promising for clinical application. We found that the scanner sensitivity is of the order of 10<sup>-5</sup> coincidences per primary proton, while the precision of the range assessment for both SPB and SOBP irradiation plans was found below 1 mm. Among the scanners with the same number of detector modules, the best results are found for the triple-layer dual-head geometry. The results indicate that the double-layer cylindrical and triple-layer double-head configurations are the most promising for the clinical application, CONCLUSIONS:: We performed simulation studies demonstrating that the feasibility of the J-PET detector for PET-based proton beam therapy range monitoring is possible with reasonable sensitivity and precision enabling its pre-clinical tests in the clinical proton therapy environment. Considering the sensitivity, precision and cost-effectiveness, the double-layer cylindrical and triple-layer dual-head J-PET geometry configurations seem promising for future clinical application.
radiology, nuclear medicine & medical imaging