Laser assisted vascular welding with real time temperature control

R B Stewart,A Benbrahim,G M LaMuraglia,M Rosenberg,G J L'Italien,W M Abbott,R T Kung
DOI: https://doi.org/10.1002/(SICI)1096-9101(1996)19:1<9::AID-LSM2>3.0.CO;2-W
Abstract:Background and objective: Previous studies in laser assisted vascular welding have been limited by the lack of a reliable end point for tissue fusion. As a means of improving the reproductibility of laser assisted repairs, a system incorporating real time temperature monitoring and closed loop feedback was used. Study design/materials and methods: The system consisted of a direct view infrared thermometer for monitoring the laser heated spot, a 1.9 microns diode laser, and a microprocessor for data acquisition and feedback control of the laser power to maintain a constant tissue temperature. Rat aortas were welded under constant surface temperature conditions. Results: In vivo temperature stability of +/- 2 degrees C was achieved over a temperature range of 70-90 degrees C pertinent to welding small vessels. When welds were completed using the feedback system to maintain the tissue temperature at 80 degrees C, the acute success rate was 100% and the burst pressure was 290 +/- 70 mmHg. Conclusion: These studies demonstrate that the use of real time monitoring and feedback control results in improved consistency for vascular tissue welding.