A prototype scintillating fibre beam profile monitor for Ion Therapy beams

B.D. Leverington,M. Dziewiecki,L. Renner,R. Runze
DOI: https://doi.org/10.1088/1748-0221/13/05/P05030
2019-10-03
Journal of Instrumentation
Abstract:A prototype plastic scintillating fibre based beam profile monitor was tested at the Heidelberg Ion Therapy Centre/ Heidelberg Ionenstrahl Therapiezentrum (HIT) in 2016 to determine its beam property reconstruction performance and the feasibility of further developing an expanded system. At HIT protons, helium, carbon, and oxygen ions are available for therapy and experiments. The beam can be scanned in two dimensions using fast deflection magnets. A tracking system is used to monitor beam position and to adjust scanning magnet currents online. A new detector system with a finer granularity and without the drift time delay of the current MWPC system with a similar amount of material along the beamline would prove valuable in patient treatment. The sensitive detector components in the tested prototype detector are double-clad Kuraray SCSF-78MJ scintillating fibres with a diameter of 0.250 mm wound as a thin multi-layer ribbon. The scintillation light is detected at the e...
instruments & instrumentation
What problem does this paper attempt to address?