A fast plastic scintillator for low intensity proton beam monitoring

Adélie André,Christophe Hoarau,Yannick Boursier,Afef Cherni,Mathieu Dupont,Laurent Gallin Martel,Marie-Laure Gallin Martel,Alicia Garnier,Joel Hérault,Johan-Petter Hofverberg,Pavel Kavrigin,Christian Morel,Jean-François Muraz,Maxime Pinson,Giovanni Tripodo,Daniel Maneval,Sara Marcatili
2024-11-12
Abstract:In the context of particle therapy monitoring, we are developing a gamma-ray detector to determine the ion range in vivo from the measurement of particle time-of-flight. For this application, a beam monitor capable to tag in time the incident ion with a time resolution below 235 ps FWHM (100 ps rms) is required to provide a start signal for the acquisition. We have therefore developed a dedicated detector based on a fast organic scintillator (EJ-204) of 25x25x1 mm3 coupled to four SiPM strips that allow measuring the particle incident position by scintillation light sharing. The prototype was characterised with single protons of energies between 63 and 225 MeV at the MEDICYC and ProteusONE facilities of the Antoine Lacassagne proton therapy centre in Nice. We obtained a time resolution of 120 ps FWHM at 63 MeV, and a spatial resolution of ~2 mm rms for single particles. Two identical detectors also allowed to measure the MEDICYC proton energy with 0.3% accuracy.
Instrumentation and Detectors,Medical Physics
What problem does this paper attempt to address?
The problem that this paper attempts to solve is to develop a fast plastic scintillator detector capable of accurately measuring the time of incident ions in particle therapy monitoring. Specifically, in order to achieve proton beam monitoring and determine the range of ions in the body, a detector that can mark incident ions in time is required, and its time resolution needs to be lower than 235 picoseconds (FWHM) to provide a start - up signal for data acquisition. ### Detailed Interpretation: 1. **Background Problem**: - Particle therapy (PT) has higher dose accuracy compared to traditional X - ray radiotherapy and can reduce the irradiation of healthy tissues. However, the accuracy of PT may be limited due to factors such as patient mispositioning, anatomical structure changes, and errors in determining tissue stopping power from X - ray images. - In order to reduce these uncertainties and improve the specificity of the technology, an online measurement of the particle range with millimeter - level accuracy is required. 2. **Research Objectives**: - Develop a gamma - ray detector based on a fast organic scintillator (EJ - 204) for determining the range of ions in the body by measuring the particle time - of - flight (TOF). - The detector must have the following characteristics: - A time resolution lower than 235 ps FWHM (100 ps rms) to provide a start - up signal. - A detection surface large enough to cover the cross - section of a typical pencil - beam (greater than 1 cm FWHM). - Almost perfect detection efficiency. - Good energy resolution to count incident protons (at least at low intensities). - The ability to measure the incident position of the beam with millimeter - level accuracy. 3. **Solutions**: - The paper describes the design and performance evaluation of a beam monitor based on a 25 × 25 × 1 mm³ fast organic scintillator (EJ - 204) and four SiPM strips. - This monitor was tested in the MEDICYC and ProteusONE facilities, using a single proton energy range from 63 to 225 MeV. 4. **Main Results**: - At a proton energy of 63 MeV, a time resolution of 120 ps FWHM was obtained. - In the clinically relevant energy range (up to 225 MeV), the time resolution remains below 235 ps FWHM. - The detection efficiency reaches 100%, and the spatial resolution is approximately 2 mm rms. Through these improvements, this detector can be used for real - time monitoring in proton therapy, ensuring high - precision particle range measurement.