Transition-Edge Sensor Microcalorimeters for X-ray Beamline Science

J.N. Ullom,W.B. Doriese,D.A. Fischer,J.W. Fowler,G.C. Hilton,C. Jaye,C.D. Reintsema,D.S. Swetz,D.R. Schmidt
DOI: https://doi.org/10.1080/08940886.2014.930806
2014-07-04
Synchrotron Radiation News
Abstract:An important and fast-growing technology in the realm of cryogenic detectors has recently been applied to beamline science: the superconducting Transition-Edge Sensor (TES). A TES consists of a thin metal film that is held in the superconducting-to-normal transition by a voltage bias. When a single X-ray photon is absorbed, the temperature and resistance of the film increase, which causes a negative-going current pulse with amplitude proportional to the deposited energy. The thin-film thermometer is typically coupled to a photon absorber composed of a high atomic number element such as bismuth or gold, so efficiencies near unity are possible for X-rays up to 10 keV. TESs are an example of a broader class of sensors called microcalorimeters that measure energetic events through a change in temperature. The energy resolution of a well-designed microcalorimeter is set by thermodynamic power fluctuations between the sensor and the local heat bath as well as by additional noise terms such as the Johnson noise in the sensing element. The energy resolution ΔE of a microcalorimeter is proportional to (kbT2C)1/2 where T is the sensor temperature and C is its specific heat. The temperature dependence of ΔE makes it advantageous to operate TESs at the lowest temperatures that can be conveniently reached, typically 0.1 K. For X-ray measurements between 1–10 keV, the dependence of ΔE on C limits the C of individual sensors to about 10−12 J/K, corresponding to device areas near 0.1 mm2 [1].
What problem does this paper attempt to address?