Calibration and Testing of Small High-Resolution Transition Edge Sensor Microcalorimeters with Optical Photons

Felix T Jaeckel,C V Ambarish,Haoran Dai,Shukai Liu,Dan McCammon,Mari McPheron,Kari L Nelms,Avirup Roy,Haley R Stueber,Simon R Bandler,James A Chervenak,Kazuhiro Sakai,Stephen J Smith
DOI: https://doi.org/10.1109/tasc.2021.3053506
2021-01-21
Abstract:Pulses of narrow line-width optical photons can be used to calibrate and test sub-2 eV full-width at halfmaximum (FWHM) energy resolution transition-edge sensor (TES) microcalorimeters at low energies (< 1 keV), where it is very challenging to obtain X-ray calibration lines comparable to (or narrower than) the detector resolution. This scheme depends on the ability to resolve the number of 3 eV photons in each pulse, which we have recently demonstrated up to photon numbers of about 300. At LTD-18 we showed preliminary results obtained with this technique on a 0.25 eV baseline resolution TES microcalorimeter designed for the ultra-high-resolution subarray of the Lynx mission. The line-shape was well described by a simple Gaussian. However, the difficulty of delivering photons to the small 46 μm square absorbers resulted in a large thermal crosstalk signal, whose random nature is expected to rapidly degrade the observed energy resolution towards higher photon numbers/energies. We have since improved the coupling between the optical fiber and the TES absorber and report here our current results.
What problem does this paper attempt to address?