High Performance Liquid Chromatography Determination of Pcmx In Blood Plasma Using Electrochemical Detection

A. Dorantes,S. Stavchansky
DOI: https://doi.org/10.1080/00032718908051616
1989-06-01
Analytical Letters
Abstract:A rapid and specific high performance liquid chromatographic (HPLC) procedure has been developed for the determination of p-chloro-mxylenol (PCMX) in blood plasma. the method is based on the extraction of PCMX from plasma with benzene in the presence of a known amount of dichloro-m-xylenol (internal standard). the benzene extract is evaporated to dryness and the residue dissolved in 200 μl of mobil phase. the HPLC system consisted of a reversed-phase column and a 65% methanol:35% ammonium carbonate 0.05% solution as a mobile phase. an electrochemical (EC) detector/glassy carbon electrode set a potential of +0.9V versus Ag/AgCl/3M NaCl is used to monitor the drug. the recovery of PCMX is approximately 98%, the limit of quantitation is 2 ng/ml of plasma for the HPLC-EC procedure. the coefficient of variation is 5.1% over the range of 10-1000 ng/ml of plasma. Data are presented to illustrate the practicality of this method for evaluation of PCMX plasma levels after a single intravenous administration of 500 mg dose of PCMX to five mongrel dogs.
chemistry, analytical
What problem does this paper attempt to address?