Determination of Helicid in Dog Plasma by RP-HPLC

Zheng Min
DOI: https://doi.org/10.3969/j.issn.1006-4931.2010.07.009
2010-01-01
Abstract:Objective To establish a RP-HPLC method to detect the concentration of helicid in dog plasma.Methods The mobile phase consisted of 0.05 mol/L sodium dihydrogen phosphate(pH 3.5)-methanol(77 :13) was used.After pretreated by a cetonitrile,the plasma samples were analyzed by RP-HPLC.Results The linear rang e of helicid was 0.025-5.000 μg/mL.And the recovery rate was 88.16%-97.56%.Both the within-day and between-day RSD were less than 12.38%.The minimum detective limit was 10 ng/mL.Conclusion This RP-HPLC method is accurate,stable and sensitive,which can be applied to the pharmacokinetic study of helicid.