Determination catecholamines in plasma and urine by RP-HPLC-ECD

王忠永,陶志华,杨建荣,戴菁,周武,陈晓东
DOI: https://doi.org/10.3969/j.issn.1001-5949.2003.08.005
2003-01-01
Abstract:Objectives To establish a method to detect catecholamines in plasma and 24h urine by RP-HPLC-ECD. MethodsThe CAs in plasma and 24h urine were absorbed by Al 2O 3,then measured using DHBA as the interal standard.Results Linear calibration curves for NE,E,DA were measured within in a range of 0.156-100 ng/ml,and had a good correlation(r=0.9998- 0.9999),the recovery rate was 63.2%-111.3% and the intraassay precision(CV) was 1.23%-19.21% while interassay CV was 2.01-19.52%.Conclusions This was a simple,accurate,high specific method to detect CAs in the clinical samples and suggested to be a sensitive,reliable method for the dianosis of phechromocytoma.