An HPLC–MS/MS Method for Quantitation of Gly-MCA in Mouse Plasma: Application to a Pharmacokinetic Study

Jia Liu,Guan Lian,Ting Wang,Yuanheng Ma,Junto Zhou,Changtao Jiang,Yuxin Yin
DOI: https://doi.org/10.1016/j.jpba.2017.07.020
IF: 3.571
2017-01-01
Journal of Pharmaceutical and Biomedical Analysis
Abstract:A simple, sensitive and rapid method using high performance liquid chromatography-tandem mass spectrometry (HPLC-MS/MS) was developed and validated for quantification of glycine-β-muricholic acid (Gly-MCA) in mouse plasma for the first time. Plasma samples were pretreated with protein precipitation. The analyte and internal standard were separated on a Shimadzu Shim-pack XR-ODS column (4.6×50mm, 2.2μm) using 0.1% formic acid-water-methanol as mobile phase, with a runtime of 5min. Detection was performed using negative ion electrospray tandem mass spectrometry via multiple reaction monitoring (MRM) scan mode. The linear range was 5ng-2μg/ml (correlation coefficient >0.995) for Gly-MCA with a lower limit of quantitation of 5ng/ml. The intra-day and inter-day precision were less than 9.2% for the analyte and accuracy was from 0.4% to 7.0%. This analytical method was then successfully applied to a pharmacokinetic study of Gly-MCA following oral administration and intraperitoneal injection in mice.