Quantification of total and free mycophenolic acid in human plasma by liquid chromatography with fluorescence detection.

Jie Shen,Zheng Jiao,Yun-Qiu Yu,Ming Zhang,Ming-Kang Zhong
DOI: https://doi.org/10.1016/j.jchromb.2004.12.005
IF: 3.318
2005-01-01
Journal of Chromatography B
Abstract:A simple high-performance liquid chromatographic (HPLC) method was developed for the assay of total and free mycophenolic acid (MPA) in human plasma. Prior to analysis, total mycophenolic acid was extracted by protein precipitation and free drug was isolated from plasma samples using ultrafiltration. The extracts were injected onto a Kromasil C8 column at 30°C with excitation and emission wavelengths set at 342 and 425nm, respectively. The mobile phase was consisted of acetonitrile-32mM glycine buffer, pH 9.2 (20:80, v/v), at a flow rate of 1.0ml/min. The method was found to be linear over the concentration range investigated, 0.05–40mg/l for total mycophenolic acid (r>0.999) and 5–1000μg/l (r>0.99) for free drug. The percentage error of the analytical method was below 10.9%. The intra- and inter-day reproducibility was adequate with the coefficients of variation of 8.28% or below. The run time were 4 and 6min for free and total MPA, respectively. The method thus can be effectively applied to measure mycophenolic acid concentrations in clinical samples.
What problem does this paper attempt to address?