A 97.3db SNR Bioimpedance AFE with −84db THD Segmented-${\Delta\Sigma \mathrm{m}}$ Sinusoidal Current Generator and Passing-Through Instrumentation Amplifier

Qinjing Pan,Qi Luo,Tianxiang Qu,Liheng Liu,Xiao Li,Min Chen,Zhiliang Hong,Jiawei Xu
DOI: https://doi.org/10.1109/vlsitechnologyandcir46783.2024.10631538
2024-01-01
Abstract:This paper presents a high-precision bioimpedance (BioZ) analog front-end (AFE) to detect changes in tissue electrical properties as prognostic indicators for heart failure, cancer, and chronic pulmonary disease. To increase the BioZ measurement accuracy, a low-distortion sinusoidal current generator (CG) is proposed. It utilizes a lookup table (LUT) with segmented delta-sigma modulator ( $\Delta\Sigma \mathrm{M}$ ) for efficient bits extension of the IDAC, and a current amplifier (CA) to isolate the IDAC from dynamic load (e.g., BioZ, electrode impedance, and parasitic capacitors) without nonlinearity penalty. The BioZ readout proposes a pass-through IA (PT-IA) to optimize the noise performance for large and complex BioZ. Manufactured in a $0.18\mu \mathrm{m}$ CMOS process, the overall AFE, including the CG and readout, achieves −84dB THD, 97.3dB SNR ( $\text{BW}=4\text{Hz}$ ), and $0.57\mathrm{m}\Omega/\sqrt{}\text{Hz}$ sensitivity at 1Hz, while the CG and readout consume $86.7-196.5\mu \mathrm{W}$ and $54.8\mu \mathrm{W}$ , respectively. Impedance cardiography (ICG) has been successfully demonstrated with dry electrodes.
What problem does this paper attempt to address?