A 0.5-Mω/√hz Dry-Electrode Bioimpedance Interface with Current Mismatch Cancellation and Input Impedance of 100 MΩ at 50 Khz.

Qinjing Pan,Tianxiang Qu,Biao Tang,Fei Shan,Zhiliang Hong,Jiawei Xu
DOI: https://doi.org/10.1109/jssc.2022.3199492
IF: 5.4
2023-01-01
IEEE Journal of Solid-State Circuits
Abstract:This article describes a high-input-impedance, low-noise bioimpedance (BioZ) sensor interface IC for small-area dry-electrode cardio-respiratory signals monitoring. To facilitate high-precision BioZ sensing with high-impedance dry electrodes, the IC utilizes three key techniques as follows: 1) a bias control loop (BCL) to eliminate the excitation current mismatch, reducing the voltage fluctuation on high-impedance input; 2) a quiet-chopping current feedback instrumentation amplifier (QC-CFIA) to mitigate the input-signal-dependent noise; and 3) a full pre-charge (FPC) technique to cancel the input parasitic capacitance for impedance boosting. Manufactured in a 0.18- $\mu \text{m}$ CMOS process, the BioZ prototype IC occupies an area of 0.4 mm2 while consuming 15.8- and 14.4–128.4- $\mu \text{W}$ current, respectively, from the amplifier and the excitation current generator (CG). With these proposed techniques, this IC achieves a high input impedance of 100 $\text{M}\Omega $ at 50 kHz, 0.5- $\text{m}\Omega /\surd $ Hz sensitivity at 1 Hz, and a 106-dB signal-to-noise ratio (SNR). A gel-free respiration and impedance cardiography (ICG) recording has been successfully demonstrated on the human body with four 0.45-cm2 dry electrodes.
What problem does this paper attempt to address?