A 4.43 TΩ Z<inf>IN</inf> 0.0128 mm<sup>2</sup> Cascaded Instrumentation Amplifier with Input-biased Pseudo Resistor for Implantable Brain Machine Interfaces

Tao Tang,Zhengjie Xu,Yina Wei,Linqing Feng,Kedi Xu,Dongping Yang,Cheney Chambers,Shenqi Qu
DOI: https://doi.org/10.1109/EMBC40787.2023.10340422
2023-01-01
Abstract:A cascaded instrumentation amplifier (CaIA) with input-biased pseudo resistors (IBPR) is presented for implantable brain machine interfaces (BMI). The gain distribution of two-stage cascaded amplifiers, instead of a single-stage amplifier, helps to achieve an input impedance of 4.43TΩ at 100Hz, and maintain the small active area (0.0128 mm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> ). The input-biased pseudo resistors contribute to a much lower high-pass corner (f <inf xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">HP</inf> =0.00011Hz) compared with the conventional structure, the input-referred noise is only 3.836μVrms integrated from 0.5Hz to 10kHz with 0.98μW power consumption.Clinical Relevance— This establishes an area-efficient amplifier design with ultra-high input impedance (4.43TΩ at 100Hz) and hyper-low high-pass corner frequency (f <inf xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">HP</inf> =0.00011Hz), which is suitable for long-term monitoring of neural activities (including slow oscillations) in implantable brain-machine interfaces.
What problem does this paper attempt to address?